Recall Z-0062-05 ยท FRN Pump, Infusion
Infusion Dynamics, A Div. of Zoll Med Corpinfusion Dynamics
Infusion Dynamics Power Infuser pump. Model number M100B-3A
| Date | Nov 2, 2004 (initiated Sep 21, 2004) |
| Status | Terminated |
| Reason | Front Panel may be cracked on some units. If cracked fluid can enter the infuser and short the start/stop button. Shorting would cause the infuser to stay in the open position. |
| Root cause | Other |
| Action | The recalling firm issued a recall letter dated 9/20/04 to their customers informing them of the problem and the need to return the units. |
| Quantity | 255 units |
| Distribution | The products were shipped to dealers in TX, MO, and VT; medical facilities in CA, NV, and NY. The product was also shipped to government accounts in PA and Germany. |
| FDA record | Open on fda.gov |