Recall Z-0037-05 ยท JOS Electrode, Electrosurgical

3M Company / Medical Division

NOVAPLUS Electrosurgical Grounding Pads, Split with Cord Catalog No. 7179V

DateOct 22, 2004 (initiated Sep 30, 2004)
StatusTerminated
ReasonImproper wire placement from the patient plate to the ESU connector plug results in an intermittent displacement connection.
Root causeOther
ActionA recall notification letter dated 09/30/2004 was sent to Hospitals and Health Care Distributors instructing consignees to immedately remove all effective product from lot 2007-06 HP.
Quantity384 cases containing 40 pads each for a total of 15,360 pads
DistributionThe subject product lot was distributed to the following key 3M distributors: Owens & Minor, Cardinal Health, McKesson, Professional Health Supply, and the Burrows Company.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K8333643M SCOTCHPLATE II DISPERSIVE ELECTRODE