Recall Z-1489-04 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Medtronic Inc

Medtronic DLP VAD Venous Cannula for Ventricular Assist, Models 95036 and CB95036.

DateSep 29, 2004 (initiated Sep 15, 2004)
StatusTerminated
ReasonSpiral tip from DLP VAD cannula may be loose and potentially detach during use.
Root causeOther
ActionA notification letter dated September 23, 2004 was sent by certified mail to US customers. The letter cautions the customer about the possibility of tip detachment and requests the impacted product to be returned to Medtronic for replacement. Customers are requested to complete a Field Corrective Action Certificate acknowledging the receipt of this information.
Quantity753 units, 203 distributed in US to 24 customers.
DistributionNationwide throughout the United States and worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K845045VENOUS RETURN CATHETER SS-1
Dlp, Inc., Apr 3, 1985
K891687EXTRACORPOREAL CIRCUIT WITH BIO-ACTIVE SURFACE
Medtronic Blood Systems, Inc., May 31, 1989