Recall Z-1485-04 ยท CBK Ventilator, Continuous, Facility Use

Cardinal Health 203, Inc

The LTV Series ventilators provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. This is a prescription medical device intended for use by personnel under the direction of a physician. The ventilator is suitable for use in institutional, home and transport settings.

DateSep 28, 2004 (initiated Sep 8, 2004)
StatusTerminated
ReasonVentilators have not consistently switched to internal battery operation when the external DC power source has become inadequate to supply stable power. This situation results in loss of ventilation to the patient.
Root causeOther
ActionA letter titled "Important Safety Information", dated September 7, 2004, was issued to all affected customers alerting them to the possibility of ventilator failure and providing instructions for recognizing and responding to the situation.
Quantity10,299 ventilators
DistributionNationwide and worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002881VENTILATOR, CONTINUOUS (RESPIRATOR), MODEL LTV 1000
Pulmonetic Systems, Inc., Oct 11, 2000
K010608MODIFICATION TO LAP TOP VENTILATOR (LTV)
Pulmonetic Systems, Inc., Mar 20, 2001
K032226LTV-1000 VENTILATOR
Pulmonetic Systems, Inc., Oct 16, 2003
K981371LTV 1000 VENTILATOR
Pulmonetic Systems, Inc., Oct 30, 1998
K984056LTV 1000
Pulmonetic Systems, Inc., Apr 12, 1999