Recall Z-1474-04 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories HPD/ADD/GPRD

XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin in Phosphate buffer, with Sodium Azide as a preservative; 950 mL bottle, 4 bottles per carton; Abbott Laboratories, Abbott Park, IL 60064 USA

DateSep 18, 2004 (initiated Sep 1, 2004)
StatusTerminated
ReasonThe XSYSTEMS Dilution Buffer when used with TDx/TDxFLx Benzodiazepines assay can cause shifts in control values and patient results.
Root causeOther
ActionAbbott sent Device Correction letters dated 9/1/04 to all customers receiving the TDx/TDxFLx Benzodiazepines assay, list 9674, informing them of the shift in control values and patient results when used with the XSYSTEMS Dilution Buffer, list 9519-02. The letter gave examples of the effects observed in the benzodiazepines assay performance, and listed the following actions to be taken until Abbott completes their investigation: * During the use of a single lot of buffer, recalibrate the TDx/TDxFLx Benzodiazepines assay every 14 days. * Perform a new calibration run on the TDx/TDxFLx Benzodi
Quantity22,442 cartons
DistributionNationwide and internationally to Argentina, Australia, Brazil, Chile, Colombia, Costa Rica, El Salvador, Germany, Guatemala, Honduras, Hong Kong, Mexico, Panama, Peru, Puerto Rico, Singapore, South Korea, Taiwan, Thailand, the United Kingdom, Uruguay and Venezuela
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.