Recall Z-1433-04 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc

COULTER LH 500 Hematology Analyzer part numbers: 178832, 178833, and 178834 software versions 1A and 1A2. The LH 500 Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter For In Vitro Diagnostic Use in clinical laboratories. The LH %500 Analyzer also provides a semi-automated reticulocyte analysis.

DateSep 2, 2004 (initiated Aug 12, 2004)
StatusTerminated
ReasonBeckman Coulter has confirmed an issue associated with the predilute mode of the COULTER LH 500 Hematology Analyzer. The workstation may display, transmit, and / or print an erroneous result or an incorrect dilution factor with a predilute sample result.
Root causeOther
ActionCustomers were notified via U.S. mail by a Product Corrective Action Letter (PCA) dated August 12, 2004. The letter includes a description of what conditions need to be met for the error to occur and the proper steps to prevent the issue from occurring.
Quantity222
DistributionNationwide & Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032000COULTER LH 500 HEMATOLOGY ANALYZER
Beckman Coulter, Inc., Aug 13, 2003