ClearanceBrief
Recall Z-1400-04 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

Magnetom Sonata MRI System, Model 7388148

DateAug 24, 2004 (initiated Jul 28, 2004)
StatusTerminated
ReasonAn input error can be made with Magnetom MRI software. Software allows users to manually change Transmit Reference Voltage in certain windows
Root causeOther
ActionThe recalling firm issued a Customer Safety Advisory letter via certified mail to their customers per Update Instructions MR028/04/S. The letter informs customers of the potential issue and provides instructions to avoid its occurence.
Quantity2 units
DistributionThe products were shipped to 39 medical facilities in CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, MA, MI, MN, MO, NC, NY, OH, OR, PA, TX, UT, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993731SONATA GRADIENTS/MAGNETOM SYMPHONY SYSTEM
Siemens Medical Solutions USA, Inc., Dec 23, 1999