Recall Z-1413-04 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Conmed Corporation

ConMed Stealth Coated Laparoscopic Electrodes with 4mm pin (60-5158 series) as follows: Cat. #60-5158-034 - Spatula, 5 mm x 32 cm; Cat #60-5158-432 - ''L'' Hook 5 mm x 32 cm; Cat #60-5158-934 - Needle 5mm x 32 cm; Firm on label: ConMed Corporation, Utica, New York 13502.

DateAug 28, 2004 (initiated Aug 9, 2004)
StatusTerminated
ReasonInsulation may pull away from the hub of the electrode exposing the conductive shaft below. This condition results in potential for electrical shock.
Root causeOther
ActionLetters dated 8/9/04 via Priority Mail with instructions to return product and subrecall if further distributed.
Quantity1,005 units
DistributionThere are 127 distributors, hospitals and sales representatives in the U.S. There are 21 foreign consignees in Australia, Brazil, Canada, Costa Rica, Ecuador, Finland, India, Israel, Italy, Norway, Peru, Saudi Arabia, Spain, Turkey, United Kingdom, and Venezuela.