Recall Z-1394-04 ยท IYX Camera, Scintillation (Gamma)

General Electric Med Systems

SMV FX-80 Nuclear Camera

DateAug 24, 2004 (initiated Jul 30, 2004)
StatusTerminated
ReasonThe detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.
Root causeOther
ActionThe detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector''s radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become lose, fail, or back out.
Quantity41
DistributionInside the US. Manufactured and distributed from 1993 - 2000.