Recall Z-1329-04 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Ev3

Protege GPS Biliary System (6Fr 10mm x 30mm) - Product number SER6-10-30-135 (.018) (Biliary)

DateAug 7, 2004 (initiated Jul 9, 2004)
StatusTerminated
ReasonEv3 had received customer complaints indicating that the stent device could not be used with an .035' guidewire. An investigation found that 6Fr, 135 cm (.018') stent system were labeled as a 120 cm (.035') stent system and that 6Fr 120 cm (.035') were labeled as 6Fr, 135 cm (.018').
Root causeOther
ActionSend notification letter to direct accounts and an EV3 representative will contact the direct account within the next few days to arrrange for the return of all unused product.
Quantity10
DistributionEleven (11) devices were distributed to seven (7) U.S. customers. Ten (10) devices were distributed to seven (7) clinical study sites in the U.S.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033314PROTEGE GPS SELF-EXPANDING NITINOL STENT AND DELIVERY SYSTEM
Ev3, Inc., Nov 21, 2003