ClearanceBrief
Recall Z-1364-04 · MZP Assay, Hybridization And/Or Nucleic Acid Amplification For Detection Of Hepatitis C Rna,Hepatitis C Virus

Roche Molecular Systems Inc

Amplicor [3] AV-HRP, Avidin-HRP (Horseradish Peroxidase) BGG Conjugate. Kit: AmpliCap HCV MonitorTest, Export, RUO.

DateAug 19, 2004 (initiated Jul 7, 2004)
StatusTerminated
ReasonAn increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.
Root causeOther
ActionWith respect to U.S distribution, email/letter on 6/24/04 to the local safety officers and general managers stationed at the affiliated distribution center, Roche Diagnostics Corporation of Indianapolis, IN. With respect to overseas distributors, email/letter on 6/24/04 to the local safety officers and general managers stationed at the affiliated distribution centers. Roche stated that the local safety officer is responsible to communicate the advisory both to the customers/end-users and the regulatory authority of the country. On July 6, 2004, a Revised Important Product Advice Notice w
Quantity172 kits
Distribution♦ Amplicor CT/NG CT Detection Kit, Lot No.E12231, Material No.20759392018, Expiration 2/2005 , Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. ♦ Amplicor CT/NG NG Detection Kit, Lot No.E12647, Material No.20759406018, Expiration 2/2005, Shipp
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.