Recall Z-1328-04 ยท DRF Catheter, Electrode Recording, Or Probe, Electrode Recording

Biosense Webster, Inc.

Stockert 70 Radio Frequency Ablation Generator

DateAug 7, 2004 (initiated Jul 16, 2004)
StatusTerminated
ReasonGenerator Failure resulting increase of catheter temperature that could not be controlled by power reduction.
Root causeOther
ActionCustomers were instructed to suspend use of the generators with software version 1.033 until they are upgraded with version 1.034B in a letter dated 7/15/2004. Firm issued a second letter on 9/17/04 for a software error discovered in-house during the upgrade. It instructs a workaround solution to prevent unwanted activity of the catheter.
Quantityunknown, not stated.
DistributionNationwide