Recall Z-1341-04 ยท CGZ Electrode, Ion-Specific, Chloride

Abbott Laboratories, Inc

ARCHITECT c8000 Processing Module; List Number 01G06-01. Manufactured by Toshiba Corporation, Medical Systems Division, Japan. Distributed by Abbott Laboratories Diagnostic Division, Irving, Texas

DateAug 13, 2004 (initiated Jun 10, 2004)
StatusTerminated
ReasonProcess change in the manufacturing of the AC/DC driver board used on the analyzer may result in one or more of the solid state relays to fail during use.
Root causeOther
ActionThe firm initiated the recall by Technical Service Bulletin sent to affected customers. The AC/DC boards will be replaced on each analyzer by a firm representative.
Quantity171 units and 25 spare parts distributed
DistributionProduct was distrbued nationwide and to the following foreign countries: Mexico, Australia, Germany, Hong Kong, Canada, Thailand, Grand Cayman, Chile, New Zealand, Columbia, Brazil, and Singapore.