ClearanceBrief
Recall Z-1171-04 ยท MIH System, Endovascular Graft, Aortic Aneurysm Treatment

Cook, Inc.

Cook brand Zenith AAA Bifurcated main body graft with the H&L-B One-Shot Introducer System, preloaded bifurcated main body graft - aortic graft diameter 26 mm - iliac leg diameters 11 mm - main body lengths contralateral limb 103mm ipsilateral limb 133 mm - delivery system 18 FR; Catalog no. TFB-26-103.

DateJul 27, 2004 (initiated Jun 10, 2004)
StatusTerminated
ReasonIntroducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.
Root causeOther
ActionThe firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.
Quantity44
DistributionNationwide, Canada, Denmark and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.