Recall Z-1052-04 ยท NIQ Coronary Drug-Eluting Stent

Boston Scientific Scimed

TAXUS Express Paclitaxel-Eluting MONORAIL Coronary Stent System (3.5 x 32mm). On August 5, 2004 an additional 38 lots were added to this recall, the following TAXUS Express2 Paclitaxel-Eluting Coronary Stent System Part Numbers correspond to the lot numbers listed at the end of the code information below. US part# H7493897024250 H7493897032270 H7493897020300 H7493897020350 H7493897032300 H7493897

DateJul 20, 2004 (initiated Jul 1, 2004)
StatusTerminated
ReasonOn a few occasions the balloon has failed to deflate within one minute after deployment of the stent.
Root causeOther
ActionJuly 2 and 16, 2004 recall letters were sent to user accounts, such as Hospitals and clinics of the need to recall TAXUS Express2 Paclitaxel-Eluting and Express 2 Coronary Stent Systems. Recall was due to characteristics in the delivery catheters that had the potential to impede balloon deflation during a coronary angioplasty procedure, a condition known as focal neckdown. An additional 38 lots were added to the recall on August 5, 2004, for the same reason.
Quantity205,983 units plus additional 3,000 units recalled on 8/5/04.
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.