Recall Z-1314-04 ยท MUJ System, Planning, Radiation Therapy Treatment

Varian Medical Systems Inc

medical device and software systems

DateAug 7, 2004 (initiated Jun 24, 2004)
StatusTerminated
Reasona malfunction in the software of the VARiS versions 6.2.27 and 6.2.35 which is used to set up radiation therapy devices which will reset parameters to null for future treatments if a patient receives treatment on a non-varian machine and a varian machine and a non-varian field is loaded in the treatment session.
Root causeOther
ActionNotification letter to consignees sent 06/14/2004.
Quantity28 units
Distributionprodut was distributed to 28 medical facilities, 12 in the United states, 16 foreign