Recall Z-1232-04 ยท GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered
Boehringer Laboratories Inc
Boehringer Replacement Gauge Model 9071, Lot Number 200405.
| Date | Jul 20, 2004 (initiated May 19, 2004) |
| Status | Terminated |
| Reason | The suction regulator incorporates a gauge for the verification of suction levels. The particular lot of gauges used in the production of the regulators exhibits a potential to develop a leak and cause the delivered suction to be inaccurate. |
| Root cause | Other |
| Action | All direct accounts were contacted by the recalling firm''s Customer Service Representative to determine the location of the affected product. The CSR arrange for replacement proudct. |
| Quantity | 9 units |
| Distribution | The recalled product was distributed to the following states: AL, FL, MA, MI, OR, PA, TN, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.