Recall Z-1066-04 ยท GEX Powered Laser Surgical Instrument

Carl Zeiss Ophthalmic Systems Inc

Surgical powered laser instrument, used for photocoagulation of ocular tissues in the treatment of diseases of the eye.

DateJul 20, 2004 (initiated Jun 3, 2004)
StatusTerminated
ReasonThe coating of laser mirrors in a defined manufacturing lot of the Visulink 532/U devices may partially tear open and loosen from the base plate. The mirror may misdirect the laser beam to an unintended target in or on the eye. Retinal bleeding and burns may result.
Root causeOther
ActionOn June 3, 2004, the firm had contacted each of its consigness individually via phone, fax, or both.
Quantity20
DistributionThe product was distributed directly to health professionals or hospitals.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013402VISULAS 532S
Carl Zeiss, Inc., Nov 14, 2001