Recall Z-1119-04 ยท JJF Analyzer, Chemistry, Micro, For Clinical Use

Roche Molecular Systems Inc

COBAS TaqMan Analyzer

DateJul 27, 2004 (initiated May 24, 2004)
StatusTerminated
ReasonImproper seating of a fiber optic cable which may cause pixel cross-talk leading to falsely elevated test results or cases of misalignment to sample-to-result mismatch.
Root causeOther
ActionAn Urgent Product Recall letter was sent to all customers with a COBAS TaqMan 96 and 48 analyzer on 5/24/2004 by overnight delivery with delivery confirmation. On May 25, 2004, All Roche Field Service Engineers were instructed to inspect the analyzers to confirm that the cables were properly seated.
Quantity34
DistributionUnits are sold to hospitals, clinics, and clinical laboratories thoughout the US. There are 10 VA hospitals. There is one government agency, NIH, Bethesda, MD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012966ROCHE COBAS TAQMAN ANALYZER
Roche Molecular Systems, Inc., Feb 1, 2002