Recall Z-1040-04 ยท FRN Pump, Infusion

Medtronic Inc, Neurological & Spinal Division

Synchromed II infusion pump

DateJul 20, 2004 (initiated May 11, 2004)
StatusTerminated
ReasonThe opposite version serial numbers were assigned to 20mL and 40mL SynchroMed II Model 8637 pumps. This size-type indication is used by the programmer to identify the pump size and can cause a limitation in the useable fill volume on a 40 mL pump and could cause some confusion related to the size of a 20 mL pump which identifies itself as a 40 mL pump.
Root causeOther
ActionThe physicians have been notified via letter of the problem with the pump. Appointments will be made with the patients and a Medtronic representative to correct the implanted devices.
Quantity14
DistributionWithin the United States and outside the United States.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.