Recall Z-1069-04 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

NxStage Medical, Inc.

NxStage Cartridge Express. A disposable extracorporeal blood and fluid pathway and a hemo filter. Model #CAR-124, 6 units per case.

DateJul 20, 2004 (initiated Apr 14, 2004)
StatusTerminated
ReasonPinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate
Root causeOther
ActionVoluntary recall notice dated 4/14/04 sent UPS, overnight with signature required to both ICU/CCU (Acute accounts) and Chronic accounts. Firm to remove and replace product that contains a disposable component from a new supplier that has demonstrated to be less reliable resulting in pinhole leaks.
Quantity234 units
DistributionCA, IL, IN, MD, IA, MO, NJ, OH, VT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040696MODIFICATION TO NXSTAGE SYSTEM ONE
Nxstage Medical, Inc., May 26, 2004