ClearanceBrief
Recall Z-1333-04 ยท DQA Oximeter

Welch Allyn Protocol, Inc

Welch Allyn Vital Signs Monitor 5300 Series 300 Part numbers 007-0098-00 through 007-0159-00

DateAug 10, 2004 (initiated Jun 8, 2004)
StatusTerminated
ReasonPotential speaker failure could result in absence of audible annunciation of pulse tones, patient alarms and equipment alerts.
Root causeOther
ActionThe firm contacted their customers and their distributors customers via letters dated June 8, 2004 through July 9, 2004. The letters instruct their customers to verify speaker functionality. If the customers do not hear two quick beeps or there is static then they have a faulty speaker. The letters instruct them to stop using the monitor and contact Welch Allyn to have the monitor repaired. If the monitor continues to emit two quick beeps their customers are to contact Welch Allyn to have the monitor repaired as soon as it is convenient. The customers are to verify the two beeps prior to
Quantity1828 units
DistributionWelch Allyn distributes product to hospitals, physicians and distributors nationwide and internationally.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.