Recall Z-1028-04 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Medical Systems Sales & Service Region No. America

Trolley for Intera/NT MRI System

DateJul 20, 2004 (initiated May 13, 2004)
StatusTerminated
ReasonPatient and MRI tabletop may fall if the tabletop is not positioned correctly on the MRI trolley.
Root causeOther
ActionThe firm sent a letter via certified mail to all customers on May 13, 2004 instructing their customers to check the position of the tabletop to ensure it is correctly aligned on the trolley. The firm is implementing a design modification beginning May 13, 2004. A field service engineer will contact the customers to arrange for installation of the retrofit for the trolley.
Quantity57 devices
DistributionProduct was distributed to hospitals and medical centers nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.