ClearanceBrief
Recall Z-1012-04 ยท FKX System, Peritoneal, Automatic Delivery

Baxter Healthcare Renal Div

HomeChoice and Yume Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C4471, 5C4471R, 5C4474, 5C4474R, T5C4441, T5C4441R

DateJul 20, 2004 (initiated May 24, 2004)
StatusTerminated
ReasonA software defect could result in a patient experiencing an overfill with software versions 8.5 and higher. The possibility of an overfill can occur when the Initial Drain Alarm Volume is set to OFF and a change to the volume is made at the Verify I-DRAIN:XXXXML prompt.
Root causeOther
ActionUrgent Device Correction letters dated 5/24/04 were mailed to the peritoneal dialysis nurse of facilities identified as having patients who use the HomeChoice machines. The letters informed them of the potential overfill problem and requested that they review the patient list provided with the letter and verify that it was complete. The nurses were requested to communicate the information to their home patients, and visit all of their patients who use the device in the low I-Drain Alarm Volume mode to make an interim correction, permanently changing the I-Drain Alarm Volume from OFF to 0ML in
Quantity9,026 units
DistributionNationwide, Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Denmark, Ecuador, France, Germany, Guatemala, Hong Kong, India, Ireland, Italy, Korea, Malaysia, Mexico, Nicaragua, the Netherlands, Peru, the Philippines, Portugal, Singapore, South Africa, Spain, Sweden, Switzerlan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K923065BAXTER PERSONAL CYCLER AUTOMMATED PERITONEAL DIAL
Baxter Healthcare Corp, Mar 4, 1994