Recall Z-1005-04 ยท ITX Transducer, Ultrasonic, Diagnostic

Hitachi Medical Systems America Inc

Hitachi HI VISION 6500, Model #EUB-6500, Ultrasound Diagnostic Scanner Device, Software Version V04-04 DICOM ServicePack 1 and EZU-FC5W.

DateJul 20, 2004 (initiated May 7, 2004)
StatusTerminated
ReasonDue to a software anomaly, various patient images may be intermixed within a single patient folder, which may result in mis-diagnosis and delay in treatment.
Root causeOther
ActionThe firm notified their Field Service Personnel and end-users by letter on 5/7/04 and 5/25/04.
Quantity3 units
DistributionThe devices were shipped to consignees in OH, PA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013723MODIFICATION TO: EUB-6500 DIAGNOSTIC ULTRASOUND SCANNER
Hitachi Medical Corp., Feb 6, 2002