Recall Z-1046-04 ยท FSA Lift, Patient, Non-Ac-Powered

Arjo, Inc.

Alenti Lift Hygiene Chair; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model numbers CDB8003-01-US and CDB8053-01-US (with scale)

DateJul 20, 2004 (initiated Jun 8, 2004)
StatusTerminated
ReasonThere has been an increased number of tipping incidents and subsequent patient injuries.
Root causeOther
ActionSafety Advice Notice letters dated 6/8/04 were sent by certified mail to the end users, informing them of the possibility of tipping if the Alenti catches on other equipment; if the resident is not positioned correctly on the chair; if the Alenti is used with an inappropriate resident; if the resident is left unattended, particularly when the Alenti is in an elevated position; and if the safety belt is not used as a positioning belt. The end users were provided with a laminated hang tag to place on the lift with the listed safety aspects and warnings, a copy of the revised Alenti Operating an
Quantity1,035 units
DistributionNationwide and Canada