ClearanceBrief
Recall Z-0991-04 ยท MKJ Automated External Defibrillators (Non-Wearable)

Cardiac Science, Inc.

FirstSave AED G3 Automated External Defibrillator IntelliSense (Lithium) Battery, Model 9143

DateJul 20, 2004 (initiated May 4, 2004)
StatusTerminated
ReasonSome of the recalled defibrillator batteries contain an incorrect fuse which could open resulting in a defibrillator which will not work.
Root causeOther
ActionA recall letter, dated May 10, 2004, stated that a replacement battery is enclosed with the letter, and instructed consignees to install the new battery in their defibrilllator and return the old battery.
DistributionThe batteries were distributed with the defibrillators nationwide in the United States and to foreign consignees in Europe, Asia, Latin America, the Middle East and Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031987POWERHEART AED G3, MODEL 9300
Cardiac Science, Jul 30, 2003