ClearanceBrief
Recall Z-1033-04 ยท MKJ Automated External Defibrillators (Non-Wearable)

Cardiac Science, Inc.

Power Heart AED Automatic External Defibrillator, Model 9200 RD

DateJul 20, 2004 (initiated May 24, 2004)
StatusTerminated
ReasonSome of the high voltage capacitors in the recalled external defibrillators have had extended charge times resulting in self-test failures following which the defibrillator cannot be used/cannot be turn on.
Root causeOther
ActionA recall letter, dated May 24, 2004, stated that a replacement AED is enclosed with the letter, and instructed consignees to replace their AED with the new one immediately and return the old AED.
DistributionThe recalled AED units were distributed nationwide in the United States and to foreign accounts in Europe, Asia, Latin America, the Middle East, and Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011901FIRSTSAVE STAR BIPHASIC, MODEL 9210/9200
Survivalink Corp., Feb 1, 2002