Recall Z-0976-04 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Varian Medical Systems

VariSource High Dose Rate Afterloader, Models 200, 200t, and ID series. Manufactured by Varian Medical Systems

DateJul 20, 2004 (initiated May 4, 2004)
StatusTerminated
ReasonManufacturer of brachytherapy instrument issued a Customer Technical Bulletin to clarify instructions for use of the radiation treatment medical device by healthcare practitioners.
Root causeOther
ActionThe recalling firm notified end users by letter and Customer Technical Bulletin on 5/4/04. The notification informed users not to use the Emergency retract hand wheel to recover a dummy wire and to reinforce the instructions already in the procedures that in the event of an Active wire becoming stuck out, the user should not turn the emergency retract hand wheel in excess of 20 times, or until the wire has parked and the radiation alarm has silenced. Advised of proper procedures for retracting the active wire and a warning concerning excessive turning of the emergency retract handwheel. The no
Quantity249
DistributionThe medical devices for correction had been distributed to medical facilities nationwide and internationally.