ClearanceBrief
Recall Z-0998-04 ยท JQW Station, Pipetting And Diluting, For Clinical Use

Roche Diagnostics Corp.

Tecan clinical workstation; Tecan Genesis 150 back end, Roche catalog number 03585352001.

DateJul 20, 2004 (initiated Mar 5, 2004)
StatusTerminated
ReasonIf the files are not periodically purged from the TCW computer, it may result in patient identification/sample results mismatches.
Root causeOther
ActionRoche sent each of their consignees revised Tecan database maintenance instruction via bulletin 04-042 dated 3/5/04.
Quantity11
DistributionNationwide. California, Connecticut, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, New Jersey, North Carolina, Ohio, Pennsylvania, Tennessee, Texas and Wisconsin.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.