Recall Z-0997-04 ยท JQW Station, Pipetting And Diluting, For Clinical Use

Roche Diagnostics Corp.

Tecan clinical workstation; Tecan Genesis 150/8 TCW front end COBAS (FEC), Roche catalog number 0360053001.

DateJul 20, 2004 (initiated Mar 5, 2004)
StatusTerminated
ReasonIf the files are not periodically purged from the TCW computer, it may result in patient identification/sample results mismatches.
Root causeOther
ActionRoche sent each of their consignees revised Tecan database maintenance instruction via bulletin 04-042 dated 3/5/04.
Quantity8
DistributionNationwide. California, Connecticut, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, New Jersey, North Carolina, Ohio, Pennsylvania, Tennessee, Texas and Wisconsin.