ClearanceBrief
Recall Z-0950-04 ยท HNN Knife, Ophthalmic

Surgical Specialties Puerto Rico, Inc.

Stylus Brand Angled Knives - 52-2912, Slit Knife, parallel sides, bevel up, angled, full handle, 3.0mm 52-3212, Slit Knife, parallel sides, bevel up, angled, full handle, 3.2mm

DateJul 20, 2004 (initiated Mar 3, 2004)
StatusTerminated
ReasonIncorrect placement of the blade in the foam of product results in blade poking through packaging. Compromised sterility and injury to healthcare worker.
Root causeOther
ActionCustomers were notified by e-mail and fax of the recall during the period of March 3, 2004 to April 1, 2004. Customers were requested to return the product to Surgical Specialties Corporation, Reading, PA 19606.
DistributionDistribution was nationwide and internationally
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K822112OPHTHALMIC KNIFE