Recall Z-0953-04 ยท GFF Bur, Surgical, General & Plastic Surgery

Medtronic Xomed, Inc.

Product is a surgical nasal/sinus surgical device labeled as a RAD Frontal Finesse Bur, Reference Number 18-83070

DateJul 20, 2004 (initiated Jan 29, 2004)
StatusTerminated
ReasonThe curve in the shaft of the bur was only 15 degrees not the labeled 40 degrees.
Root causeOther
ActionA Recall notification letter was mailed to each hospital account on January 29, 2004 for each type of device they received which was under recall explaining the error and asking the accounts to remove recalled product from inventory, fax them the response sheet and to return the products.
Quantity8 boxes 3 burs per box
DistributionProducts were distributed to 23 hospitals and surgical centers in NC, NB, IN, UT, TX, NY, WI, MO, TN, ID, OK, FL, MA, WY, NJ, LA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.