Recall Z-1054-04 ยท NEP System/Device, Pharmacy Compounding

Baxter Healthcare Corp.

Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA; product code 2M8290 - 1.0 mL accuracy, and product code 2M8290L - less than 1.0 mL accuracy

DateJul 20, 2004 (initiated Apr 26, 2004)
StatusTerminated
ReasonThe Micromix compounder may have ingredient delivery outside of the stated accuracy limits if the Vlier pin is not properly adjusted
Root causeOther
ActionBaxter began telephoning their customers on 4/26/04, advising them that an improperly adjusted Vlier pin could result in an ingredient delivery outside of stated accuracy limits. The accounts were requested to immediately stop using the Micromix Compounder and return the device to Baxter, and to evaluate all TPN formualtions that have been compounded but not administered to determine if any ingredients were delivered at a programmed volume of less than 1 mL. The account was advised to discard those solutions and recompound them using manual techniques. Follow-up recall letters were faxed to
Quantity500 devices
DistributionNationwide and internationally through Baxter subsidiaries to the United Kingdom, Germany, Chile, Canada, Brazil, El Salvador, Taiwan and Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K955622AUTOMIX PLUS, 3+3, MICROMIX COMPOUNDING SYSTEM;
Baxter Healthcare Corp, May 6, 1997