ClearanceBrief
Recall Z-0929-04 ยท BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Hudson Respiratory Care Inc

Concha IV Plus Heated Humidifier

DateJul 20, 2004 (initiated Apr 20, 2004)
StatusTerminated
ReasonSoftware malfunction. The humidifier did not retain the clinician selected default settings for temperature gradient, after the unit had been placed in the "pause" mode.
Root causeOther
ActionFirm will call hospitals to arrange for a visit to update the software.
Quantity1049
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.