Recall Z-0881-04 ยท FRN Pump, Infusion
Baxter Healthcare Corp.
Colleague CX Single Channel Volumetric Infusion Pumps, product code 2M8161; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in Singapore; color display screen
| Date | Jul 20, 2004 (initiated Apr 13, 2004) |
| Status | Terminated |
| Reason | Ink deterioration of the letters, numbers and symbols of the programming and pump channel keypads, which may impact the legibility of the keys over time. |
| Root cause | Nonconforming Material/Component |
| Action | Baxter sent urgent device correction letters dated 4/13/04 to the direct accounts to the attention of the Director of Biomedical Engineering on the same date. The letters informed the accounts of the ink deterioration of the letters, numbers and symbols on the Colleague Volumetric Infusion Pump programming and pump channel keypads, and informed them that over time such ink deterioration may impact legibility of the keys. The letters provided a listing of model numbers and serial numbers of pumps at their facility which may have potentially affected keypads. The accounts were given a sample |
| Distribution | Nationwide, and internationally to Australia, the Bahamas, Belgium, Brazil, Canada, Colombia, El Salvador, Guatemala, Honduras, Israel, Lebanon, Mexico, New Zealand, Panama, the Philippines, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates and the United Kingdom. |
| FDA record | Open on fda.gov |