ClearanceBrief
Recall Z-0866-04 ยท LJT Port & Catheter, Implanted, Subcutaneous, Intravascular

Deltec, Inc

Port-A-Cath II Polysulfone/Titanium Venous Access System, with PolyFlow Polyurethane Catheter, 2.6mm O.D. x 1.6mm I.D., 8.5 Fr Introducer Set, REF 21-4055-24

DateJul 20, 2004 (initiated Mar 29, 2004)
StatusTerminated
ReasonSome of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization.
Root causeOther
ActionA recall letter dated March 26, 2004 gave customers the option of returning the products or using the products if the customer's procedures do not require sterility of the outer surface of the inner tray.
DistributionThe recalled products were shipped nationwide in the United States and worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K962695PORT-A-CATH II VENOUS IMPLANTABLE ACCESS SYSTEMS PRECONNECTED WITH POLYURETHANE CATHETER
Sims Deltec, Inc., Oct 1, 1996