Recall Z-0919-04 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

bioMerieux Inc

VITEK GPS-107 Gram Positive Susceptibility Card, Catalog #V4368, for in-vitro diagnostic use, 20 cards per package. The responsible firm on the label is bioMerieux, Inc., Durham, NC.

DateJul 20, 2004 (initiated Apr 20, 2004)
StatusTerminated
ReasonSome of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly.
Root causeOther
ActionThe firm issued recall letters dated 4/20/04 via certified mail requesting remaining inventories be destroyed and explaining the method which the customer can use to confirm the cards were not misreported.
Quantity1,760/20-card boxes
DistributionDistribution was nationwide to medical facilities and laboratories. Foreign distribution was made to China and Switzerland.