Recall Z-0850-04 ยท EAM Screen, Intensifying, Radiographic

General Electric Med Systems

GE Single 15 inch TV Monitor Suspension with Rail, Model 46-240485P1, Catalog number B0256EH

DateJul 20, 2004 (initiated Mar 30, 2004)
StatusTerminated
ReasonThe ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.
Root causeOther
ActionField modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals.
Quantity1236 devices
DistributionThe units were distributed nationwide in the United States and worldwide.