Recall Z-0846-04 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

AEROSET Analyzer System, Model 9D05-01, distributed by Abbott Laboratories Diagnostic Division, Irving, Texas

DateJul 20, 2004 (initiated Mar 22, 2004)
StatusTerminated
ReasonChange of level 3 error log messages to level 1 to prevent suspect assay results.
Root causeOther
ActionThe recall was initiated by letter on March 22, 2004. The letter instructed customers regarding steps to change the error levels on each AEROSET system.
Quantity656
DistributionNationwide and to the following foreign countries: Mexico, Columbia, Puerto Rico, Germany, New Zealand, Guatemala, Argentina, Dominican Republic, Hong Kong, Romania, Venezuela, Brazil, Canada, Australia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980367AEROSET SYSTEM
Abbott Laboratories, Apr 1, 1998