Recall Z-0818-04 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

ARCHTECT System RS-232 Manual, product list number 06F71-04, for the ARCHITECT i2000 Processing Module and the ARCHITECT c8000 Processing Module. The product is manufactured and distributed by Abbott Laboratories Diagnostic Division, Irving, TX 75038

DateJul 20, 2004 (initiated Mar 19, 2004)
StatusTerminated
ReasonWhen using a LIS the software system will send up a test result of up to 20 characters although the field is set to display only 15 characters per the RS-232 Manual.
Root causeOther
ActionCustomers were notified by letter on March 23, 2004 to contact their LIS vendor and to reconfigure the ASTM Field 10.1.4 to allow a maximum character length of 20 characters.
Quantity101
DistributionNationwide and to the following foreign countries: Chile, Canada, Germany, Taiwan, Australia, Signapore