Recall Z-0826-04 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Nova Biomedical Corporation

Stat Profile Critical Care Xpress (CCX) Analyzer Catalog Number: 35942

DateJul 20, 2004 (initiated Mar 30, 2004)
StatusTerminated
ReasonIncorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
Root causeOther
ActionNova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.
Quantity35 units
DistributionCA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022746STAT PROFILE CRITICAL CARE XPRESS ANALYZER
Nova Biomedical Corp., Nov 22, 2002