ClearanceBrief
Recall Z-0802-04 ยท DQR Cannula, Catheter

Guidant Corp-Cpi Division

Guidant EasyTrak 8 French Guiding Catheter, Inner Diameter .087'', 2.21 mm, Overall Lenght 54 cm, Coronary Sinus - Amplatz 6.0 Tip Shape, CS-A6, REF 7576

DateJul 20, 2004 (initiated Mar 11, 2004)
StatusTerminated
ReasonThe terminal hub component in a some of the catheters has an unintended abrupt transition between the entry port and the catheter lumen which can result in difficulty introducing or removing delivery system components during lead placement.
Root causeOther
ActionStarting on March 11, 2003, the firm''s representatives visited consignees, to whom the recalled catheters had been consigned, and retrieved the catheters.
Quantity20 catheters
DistributionNationwide in the United States.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021284EASYTRAK GUIDING CATHETER
Guidant Corp., May 2, 2002
K021455RAPIDO GUIDING CATHETER, 6F MODEL# 6776; RAPIDO GUIDING CATHETER, 8F MODEL# 6775, 6777, 6778, 6779, 6780, 6781, 6782,
Guidant Corp., Aug 2, 2002