ClearanceBrief
Recall Z-0769-04 ยท CDO Acid, Uric, Uricase (U.V.)

Abbott Laboratories Diagnostic Div

Clinical Chemistry Uric Acid; LN 7D76-20

DateJul 20, 2004 (initiated Mar 18, 2004)
StatusTerminated
ReasonUnacceptable upward drift of results in control runs within 24 hour period.
Root causeOther
ActionFirm sent recall letters 3/18/2004 suggesting destruction and/or running QC tests every eight hours instead of every 24 hours as is in labeling.
Quantity796 kits.
DistributionNationwide and Mexico, Equador, Puerto Rico, Hong Kong, Bahamas, Venezuela, Argentina, Canada, New Zealand, Columbia, Brazil, England, Trinidad.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.