ClearanceBrief
Recall Z-0912-04 ยท MUJ System, Planning, Radiation Therapy Treatment

ADAC Laboratories

Pinnacle3 Radiation Therapy Planning System, Version 6.6c Kit, Model Number 9200-0613C-ENG

DateJul 20, 2004 (initiated Mar 3, 2004)
StatusTerminated
ReasonA software defect, relating to the motorized wedge functionality when using an Elekta linear accelerator, can result in incorrect dose delivery to the patient. Also, dose delivery in the wrong location may occur when using a Elekta Leksell headframe.
Root causeOther
ActionA notification dated March 3, 2004 was sent to the customers. The notification gave recommendations for avoiding the problem and stated that updated software would be provided when it becomes available. Firm representatives made follow-up telephone calls to the consignees.
Quantity55
DistributionThe recalled products were shipped to consignees located nationwide the United States and in Australia, Canada, China, and Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.