Recall Z-0885-04 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

DWL Systems, Inc.

EZ-DOP, diagnostic ultrasound imaging system

DateJul 20, 2004 (initiated Jan 5, 2004)
StatusTerminated
ReasonA certain sequence of data entries on the keypad could cause the ultrasound medical device to display incorrect transmission power.
Root causeOther
ActionThe firm notified consignees by mail on 1/5/2004 and advised of potential for inaccurate power display when the device is actually transmitting. Consignees were requested to return the device for a software upgrade.
Quantity367 units
DistributionUnits were distributed to end users worldwide including 67 domestic consignees.