Recall Z-0729-04 ยท DQO Catheter, Intravascular, Diagnostic

Boston Scientific Scimed

Impulse 5F 145 (degree) Multipack Angiographic Catheters, Catalog Number H749163913011

DateJul 20, 2004 (initiated Feb 13, 2004)
StatusTerminated
ReasonSome of the catheters may have yellow foreign material in their pigtail tips. The foreign material in the device could cause a stroke or significant damage to organs other than the brain.
Root causeOther
ActionConsignees were sent a recall letter dated February 13, 2004. The letter requested that consignees discontinue use of the recalled catheters and return them.
Quantity11 catheters
DistributionThe product was shipped to consignees in Florida in the United States.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.