Recall Z-0735-04 ยท JJQ Colorimeter, Photometer, Spectrophotometer For Clinical Use

Biomerieux, Inc.

NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).

DateJul 20, 2004 (initiated Feb 26, 2004)
StatusTerminated
ReasonInstrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.
Root causeOther
ActionThis correction is being handled as an instrument repair (rework) by bioMerieux Field Service; therefore no communication has been sent to the customer sites. The repair involves heat sink replacement or use of insulating tape on the rear side of the keypad.
Quantity3 units
DistributionCT, MD, MT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.