Recall Z-0721-04 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

bioMerieux Inc

VITEK GPS-107Gram Positive Susceptibility Card, Catalog #V4368, for in-vitro diagnostic use, 20 cards per package. The responsible firm on the label is bioMerieux, Inc., Durham, NC.

DateJul 20, 2004 (initiated Jan 28, 2004)
StatusTerminated
ReasonSome of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly.
Root causeOther
ActionThe firm issued recall letters dated 1/28/04 via regular mail requesting remaining inventories be destroyed and explaining the method which the customer can use to confirm the cards were not misreported. A 'Second Notification' recall letter was issued dated 3/9/04.
Quantity1,988/20-card boxes
DistributionDistribution was made nationwide to medical facilities, which includes military and government consignees. Distribution was also made to foreign consignees located in Colombia, Costa Rica, and China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.