Recall Z-0726-04 ยท FNL Bed, Ac-Powered Adjustable Hospital

Hill-Rom, Inc.

Hill Rom brand Advanta bed; model P1600.

DateJul 20, 2004 (initiated Mar 8, 2004)
StatusTerminated
ReasonA potential pinch/shear point exists between the restraint strap hook located underneath the knee section and the foot rail mounting of the bed.
Root causeOther
ActionA recall letter dated March 8, 2004 was sent to each consignee advising them of the problem, and that Hill-Rom will visit the facility to modify the beds.
Quantity294
DistributionCalifornia, Delaware, Idaho, Indiana, Massachusetts, Michigan, Mississippi, New York, North Carolina, Ohio, South Dakota, Tennesee and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.